Nichipet Premium
Click to enlarge image(s)

IAncncoevastsioonr yin “Durability” and “Strength”

  • Seal ring for the Premium is tested dispensing / aspirating 600,000 times.( 2μL ‒ 1000μL )
  • Nozzle top is highly abrasion tolerant. ( 2μL - 1000μL )
  • Non corrosive ceramic is used for the plunger ( 100μL - 10mL )

IAncncoevastsioonr yin “Stability” and “Preciseness”

  • Hyper Blower System, longer second-push stroke, improves dispensing efficiency. ( 2μL,10μL,20μL )
  • Easy-Calibration function provides simple lab calibration.
  • Hand temperature doesn’t permeate through the body of the pipette which works to avoid affecting volume measurement. ( Patented )

IAncncoevastsioonr yin “Fatigue-Free” and “Comfort”

  • New round shape of the Premium mitigates fatigue and stress from heavy workload.
  • Enlarged push button makes it easy to depress.
  • White body color mitigates reflection of room light.

*All features of Nichipet EXII are available on Nichipet Premium.



Specifications

Cat. No Volume Setting Volume range(μL) Increments(μL) Usable tips
00-NPP-2 Digital Setting 0.1~2 0.002 BMT2-UT/UTR、FUT/FUTRB
00-NPP-10 0.5~10 0.01 BMT2-SS/SSR、FSS/FSSRB
00-NPP-20 2~20 0.02 BMT2-SG/SGR、FSG/FSGRB
00-NPP-100 10~100 0.1 BMT2-SG/SGR、FSG/FSGRB※
00-NPP-200 20~200 0.2 BMT2-SG/SGR、FAG/FAGRB
00-NPP-1000 100~1000 1 BMT2-LG/LGR、LE/LER、FLG/FLGRB、FLE/FLERB
00-NPP-5000 1000~5000 10 BMT2-X/XR
00-NPP-10000 1000~10000 10 BMT2-Z/ZR

AC&CV

Cat. No Volume(μL) Accuracy(%) Precision(%) Cat. No Volume(μL) Accuracy(%) Precision(%)
00-NPP-2 0.2 ±12.0※ ≦6.0※ 00-NPP-200 20 ±1.0 ≦0.5
1 ±5.0 ≦2.5 100 ±0.8 ≦0.3
2 ±3.0 ≦1.0 200 ±0.8 ≦0.2
00-NPP-10 1 ±4.0 ≦3.0 00-NPP-1000 100 ±1.0 ≦0.5
5 ±1.0 ≦1.0 500 ±0.8 ≦0.3
10 ±1.0 ≦0.5 1000 ±0.7 ≦0.2
00-NPP-20 2 ±5.0 ≦3.0 00-NPP-5000 1000 ±1.0 ≦0.3
10 ±1.0 ≦1.0 2500 ±0.8 ≦0.3
20 ±1.0 ≦0.4 5000 ±0.6 ≦0.2
00-NPP-100 10 ±2.0 ≦1.0 00-NPP-10000 1000 ±2.0 ≦0.4
50 ±1.0 ≦0.3 5000 ±0.8 ≦0.3
100 ±0.8 ≦0.3 10000 ±0.4 ≦0.2

※The AC and CV above are measured with the disposable tips described in the catalogue.

※The AC and CV at 0.2μL depend much on the sampling skill and the sampling condition where pipettes are used.